Mindray A8 - O2 Cell Door Cover Assembly, O2 Cell Calibration Valve Assembly, and ACGO Assembly

Mindray A8
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11-8
11.1.6 O2 Cell Door Cover Assembly, O2 Cell Calibration Valve
Assembly, and ACGO Assembly
No. Name P/N Test Item After
Replacement
Remarks
1
O2 cell door cover
assembly (US) (The
door can be opened.)
115-064805-00
5.2 System Check;
5.7.1 Vaporizer
Monitoring Function Test
/
O2 cell door cover
assembly (Canada)
(The door can be
opened.)
115-064804-00
5.2 System Check;
5.7.1 Vaporizer
Monitoring Function Test
/
2
O2 Cell Calibration
Valve Assembly
115-064094-00
5.2System check;
5.7.1 Vaporizer
Monitoring Function Test
Install the
O2 cell
3 ACGO assembly 115-064074-00
5.9.4 Mechanical ACGO
function test;
5.7.1 Vaporizer
Monitoring Function Test
/