7-54 
Cause Confirmation 
Solution 
the output pressure of the 
secondary pressure regulator 
with a manometer. 
is beyond 150±5 
secondary pressure 
regulator (refer to 11.1.4). 
No AA output of 
the electronic 
vaporizer 
1. Check whether the gas 
supply is sufficient. 
pressure is less than 
Connect or replace the gas 
supply. 
vaporizer alarm is 
A vaporizer alarm 
is generated. 
Refer to 7.3.6 to clear the 
alarm. 
3. Install the electronic 
vaporizer in another slot and 
try to activate it. 
Anesthetic agent is 
output normally. 
Replace the vaporizer base 
valve assembly (refer to 
Anesthetic agent is 
not output 
normally. 
(1) Replace the electronic 
vaporizer. 
(2) If the problem persists, 
replace the vaporizer base 
valve assembly (refer to 
Inaccurate 
concentration 
regulating of the 
electronic 
vaporizer 
Check whether the gas 
supply is sufficient. 
pressure is less than 
Connect or replace the gas 
supply. 
Select Setup > Service > 
Data Monitors > 
Vaporizer to access the data 
monitoring screen and check 
the drive gas pressure of 
Vaporizer 1 (or Vaporizer 
2). 
pressure is within 
150±5 kPa (22±0.7 
Replace the electronic 
vaporizer. 
pressure is beyond 
150±5 kPa (22±0.7 
Adjust or replace the 
secondary pressure 
regulator (refer to 11.1.4). 
 
7.4.10 AGSS Failures 
Cause Confirmation 
Solution 
No response from 
the float when the 
knob is adjusted 
during ventilation 
with AGSS ON 
accessories are connected 
are not connected 
Connect the AGSS 
accessories properly. 
hospital's waste gas 
scavenging system is 
waste gas 
scavenging system 
Properly connect to the 
waste gas scavenging 
system. 
Remove the AGSS 
assembly by referring to 
8.34 and check whether 
the sampling line of the 
AGSS sensor falls off. 
of the AGSS 
Reconnect the sampling 
line of the AGSS sensor. 
The sampling line 
of the AGSS 
sensor does not 
fall off. 
cable. 
(2) If the problem persists, 
replace the AGSS 
assembly and cable (refer