7-49 
Cause Confirmation 
Solution 
Check the fresh air flow 
setting. 
flow setting is 
beyond the 
Adjust the fresh air flow 
setting to be within the 
optimizer range. 
Check the zero point of the 
sensor by referring to 5.15. 
Refer to 5.15. 
Zero the sensor by referring 
to 6.3.7. 
measurement of the pressure 
sensor is accurate by 
referring to 
5.17 (device 
required) or 7.6.1.3
 (no 
Make a judgment 
by referring to 
5.17 or 7.6.1.3. 
calibration by referring to 
6.3.5. 
(2) If the problem persists, 
replace the sensor adapter 
board (refer to 11.1.16). 
Perform inspiratory valve 
diagnosis by referring to 
A diagnosis error 
occurs. 
Replace the drive gas 
assembly (refer to 11.1.17). 
Open the service door (refer 
to 8.1.1) and check whether 
the sensor adapter board and 
3-way valve sampling line 
The silicone tube 
loosens or is 
occluded. 
Reconnect the silicone tube 
and prevent it from 
occlusion. 
Remove the VE assembly by 
referring to 8.29 and check 
whether the sampling line on 
the work surface is connected 
The silicone tube 
loosens or is 
occluded. 
Reconnect the silicone tube 
and prevent it from 
occlusion. 
technical alarms are 
technical alarm is 
Refer to 7.3 to clear the 
alarm. 
The problem persists. 
Replace the drive gas 
assembly (refer to    11.1.17). 
 
7.4.6 Inaccurate PEEP Control 
Cause Confirmation 
Solution 
Inaccurate PEEP 
control 
Check whether the gas 
supply is sufficient. 
pressure is less 
than 280 kPa (40 
Connect or replace the gas 
supply. 
Check the zero point of the 
sensor by referring to 5.15. 
Refer to 5.15. 
Zero the sensor by referring 
to 6.3.7. 
measurement of the pressure 
sensor is accurate by 
referring to 
5.17 (device 
required) or 7.6.1.3
 (no 
Make a judgment 
by referring to 
5.17 or 7.6.1.3. 
Perform pressure calibration 
by referring to 6.3.5. 
Open the service door (refer 
to 8.1.1) and check whether 
the sensor adapter board and 
loosens or is 
Reconnect the silicone tube 
and prevent it from