11-29
11.5 Breathing Circuit Subsystem
11.5.1 Patient Circuit Material Package
No. Name P/N
Test Item
After Replacement
Remarks
1
Bag arm assembly
115-048600-00
5.2 System Check
/
Flexible bag arm
115-048035-00
5.2 System Check
/
2
Auto/manual valve
115-064186-00
5.2 System Check
/
3 APL valve assembly 115-064061-00
5.10.2 APL valve test
5.2 System Check
/
4
Bypass assembly
115-066853-00
5.2 System Check
/
5
Sodalime absorber canister
assembly
115-066854-00 5.2 System Check /
6
CO2 Locking assembly
115-067919-00
/
/
7
O2 cell holder assembly
115-064181-00
5.2 System Check
/
8
Patient circuit assembly
115-066852-00
5.2 System Check
/
/
Inspiration flow sensor
115-041507-00
6.3.1 Flow calibration
/
/
Expiration flow sensor
115-041508-00
6.3.1 Flow calibration
/
3
4
8
1
2
7
5
6