7-27 
 
7.3.6 Electronic Vaporizer Runtime Alarms 
Liquid Level Low 
liquid level is lower 
than 50 ml and higher 
than or equal to 20 
ml. 
1. Observe the remaining anesthetic 
agent through the glass tube. If the 
anesthetic agent is insufficient, fill 
the anesthetic agent. 
2. If the anesthetic agent is 
sufficient, reinstall the electronic 
vaporizer. 
3. If the problem persists, replace 
the electronic vaporizer. 
4. After the electronic vaporizer is 
replaced and inserted, conduct the 
power-on selftest, leak test, and 
electronic vaporizer test (refer to 
Almost Empty 
liquid level is lower 
than 20 ml. 
1. Observe the remaining anesthetic 
agent through the glass tube. If the 
anesthetic agent is insufficient, fill 
anesthetic agent. 
2. If the anesthetic agent is 
sufficient, reinstall the electronic 
vaporizer. 
3. If the problem persists, replace 
the electronic vaporizer. 
4. After the electronic vaporizer is 
replaced and inserted, conduct the 
power-on selftest, leak test, and 
electronic vaporizer test (refer to 
is Open 
Information 
2. Non-standby 
and non-active 
state: 
Information 
(first 2 min), 
Low (2 min 
later) 
3. Non-standby 
and active state: 
Information 
(first 2 min), 
Med (2 min 
open. 
1. If the vaporizer lid is open, close 
it. 
2. Reinstall the electronic vaporizer. 
3. If the problem persists, replace 
the electronic vaporizer. 
4. After the electronic vaporizer is 
replaced and inserted, conduct the 
power-on selftest, leak test, and 
electronic vaporizer test (refer to 
5.7). 
Failure 
boil-off space 
temperature too high 
(2) AA vaporizer 
boil-off space 
temperature sensor 
Rectify the fault based on the alarm 
error code by referring to 7.3.6.2.