7-28 
injection branch 
failure 
(4) AA vaporizer 
heating board failure 
(5) AA vaporizer 
drive gas branch 
failure 
(6) AA vaporizer 
AA Vaporizer 
Selftest Error 
An AA vaporizer 
self-test error occurs. 
Rectify the fault based on the alarm 
error code by referring to 7.3.6.1. 
Comm Stop 
the CPU board stops. 
1. If the status screen has a 
communication stop alarm, or 
encounters the black or blank screen 
error, rectify the fault of the status 
screen first (refer to 7.2.1). 
2. Reinstall the electronic vaporizer. 
3. Replace the electronic vaporizer 
slot. If the problem is solved, 
restart the machine. 
If the problem persists, replace the 
electronic dock adapter board (refer 
to 11.4.1). 
If the problem persists, replace the 
monitoring and controlling assembly 
(refer to 11.1.11). 
4. If the problem persists, replace 
the electronic vaporizer. 
5. After the electronic vaporizer is 
replaced and inserted, conduct the 
power-on selftest, leak test, and 
electronic vaporizer test (refer to 
Output 
Concentration 
Measure 
Abnormal 
anesthetic agent 
concentration of the 
vaporizer greatly 
deviates from the 
setting. 
1. Reinstall the electronic vaporizer. 
2. Select Setup > Service > Data 
Monitors > Vaporizer . On the 
vaporizer diagnostics screen, check 
the pressure of the vaporizer drive 
gas. If the drive gas pressure is 
beyond the range of 140-160kPa (20 
psi–23 psi), replace the electronic 
vaporizer pressure regulator (refer to 
11.1.4). 
3. If the drive gas pressure is 
normal, or the problem persists after 
the adjustment, connect the external 
AG module, and test the electronic 
vaporizer concentration according to 
section 7.8.3. Select Setup > 
Service > Data Monitors > AG, 
and check the measured values of