7-45 
connection cable between the 
backplane and the display. 
4. If the problem persists, replace the 
module rack (refer to 11.1.14). 
5. If the problem persists, replace the 
cable between the backplane and 
NMT Comm. Stop  High 
NMT communication 
stops. 
1. Re-plug the NMT module, main 
cable, sensor, and stimulation 
electrode, or plug the NMT module to 
another machine. Eliminate the error 
situations one by one. 
2. If the alarm persists, replace the 
NMT module or related component. 
3. If the problem persists, check the 
connection cable between the 
backplane and the display. 
4. If the problem persists, replace the 
module rack (refer to 11.1.14). 
5. If the problem persists, replace the 
cable between the backplane and 
NMT Init Error  High 
An NMT initialization 
error occurs. 
1. Re-plug the NMT module or restart 
the anesthesia machine. 
2. If the alarm persists, replace the 
NMT module. 
3. If the problem persists, check the 
connection cable between the 
backplane and the display. 
4. If the problem persists, replace the 
module rack (refer to 11.1.14). 
5. If the problem persists, replace the 
cable between the backplane and 
NMT Self Test Error  Low 
An NMT self-test error 
occurs. 
1. Re-plug the NMT module or restart 
the anesthesia machine. 
2. If the alarm persists, replace the 
NMT module. 
3. If the problem persists, check the 
connection cable between the 
backplane and the display. 
4. If the problem persists, replace the 
module rack (refer to 11.1.14). 
5. If the problem persists, replace the 
cable between the backplane and 
NMT Power Error  High 
disconnected, or an NMT 
communication error 
1. Check the connection between the 
main cable and the NMT module. 
2. If the alarm persists, replace the 
NMT module or main cable. 
Low 
An NMT power supply 
error occurs. 
1. Re-plug the NMT module. 
2. If the alarm persists, replace the