7-56 
Cause Confirmation 
Solution 
or HFNC module (refer to 
11.1.4). 
Failure to adjust 
the total flow or 
O2 concentration 
of the auxiliary 
O2/air or HFNC 
module 
supply is connected properly 
(O2 supply is required for 
the HFNC module), and is 
The O2 supply 
pressure is less than 
280 kPa (40 psi). 
Connect or replace the gas 
supply. 
Check whether the internal 
airway is connected 
properly. (Remove the 
auxiliary outlet by referring 
The internal 
pipeline falls off or 
leaks, with gas 
leaking sound. 
Reconnect or replace the 
pipeline. 
Reconnect the encoder cable 
by referring to 8.54 (remove 
the instrument panel by 
The parameters can 
be adjusted after 
the reconnection. 
/ 
Exchange the total flow 
encoder and O2 
concentration encoder to 
check whether the total flow 
encoder malfunctions (refer 
to 8.54). 
the original 
abnormal 
parameters can be 
adjusted, but the 
original normal 
parameters cannot 
Replace the flow encoder 
(refer to 11.1.5). 
abnormal 
parameters still 
cannot be adjusted 
Replace the auxiliary O2/air 
or HFNC module (refer to 
11.1.4). 
Auto O2 
concentration or 
flow change of 
the auxiliary 
O2/air or HFNC 
module, with 
adjustment 
failure 
Check whether the two gas 
supplies are connected 
properly and are sufficient. 
 
 
supply pressure is 
below 280 kPa (40 
Connect or replace the gas 
supply. 
Both gas supplies 
are insufficient. 
(1) Reconnect the internal 
pipeline. 
(2) If the problem persists, 
replace the auxiliary O2/air 
or HFNC module (refer to 
No gas output of 
the auxiliary 
O2/air or HFNC 
module 
Check whether the gas 
supply is sufficient. 
supply pressure is 
below 280 kPa (40 
Connect or replace the gas 
supply. 
The gas supply is 
sufficient. 
(1) Reconnect the internal 
pipeline. 
(2) If the problem persists, 
replace the auxiliary O2/air 
or HFNC module (refer to