6-24
1. Enter the standby mode and turn HFNC on.
2. Select Setup > System > Calibrate > Zero HFNC to access the screen shown below.
3. Select Begin to start zeroing. During zeroing, you can select Cancel to abort the zeroing.
4. The screen shown below is displayed if the ongoing zeroing is aborted. Select Retry to do
the zeroing again, or select Done to exit the zeroing screen.