EasyManuals Logo

Mindray A8 Service Manual

Mindray A8
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #190 background imageLoading...
Page #190 background image
6-28
6.3.5.3 Calibration Procedure
Automatic Calibration
Perform the following steps to calibrate the pressure sensor and PEEP proportional valve.
1. Ensure that the anesthesia machine is in standby mode.
2. Select Setup > Service > Calibrate > Pressure Sensor and select Automatic Calibrate to
access the screen shown below.
3. Use a communication cable to connect the calibration device and the anesthesia machine
according to the instructions on the screen. Connect the pressure sampling line to the
calibration device according to steps 2 and 4. Details are shown below:
VT Plus setup:
When VT Plus is used as the calibration device, perform the following operations on it (note:
verify the correctness of the settings before each calibration):
1) Zero Mode: Manual
Select SETUP > Settings > ENTER and set Zero Mode to Manual.
2) Serial Mode: OTIS Ctrl
Select SETUP > System > ENTER and select Serial Mode: OTIS Ctrl.
Note: Set Serial Mode before calibration. Exit the Serial Mode setup screen before you
access the (flow or pressure) measurement screen. To exit the Serial Mode setup screen,
Select SETUP > System > ENTER and select Serial Mode: NONE.
VT 650/900 setup :
When VT 650/900 is used as the calibration device, perform the following operations on it (note:
verify the correctness of the settings before each calibration):
Select Menu > Unit > Low Pressure, select cmH2O, and select OK.

Table of Contents

Other manuals for Mindray A8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A8 and is the answer not in the manual?

Mindray A8 Specifications

General IconGeneral
BrandMindray
ModelA8
CategoryMedical Equipment
LanguageEnglish

Related product manuals