6-59
7. The screen shown below is displayed upon successful calibration. Select Done to exit the
calibration screen.
8. The vaporizer should be checked by a calibrated gas tester after vaporizer injector calibration.
6.3.12.3 Common Failures and Recommended Actions
Failure
Description
Possible Cause Recommended Action
00000001 There is no drive gas.
1. Check whether the gas supply
is connected and is sufficient
(280–600 kPa (40 psi–87
psi)).
2. Check whether the cover of
the electronic vaporizer is
closed.
3. Replace the electronic
vaporizer.
4. Replace the electronic
vaporizer manifold.
00000002 The fresh gas flow is invalid. 1. Check whether the gas supply
is connected and is sufficient
(280–600 kPa (40 psi–87
psi)).
2. Check whether the electronic
flowmeter works properly.
00000004 There is no fresh gas.
1. Check whether the gas supply
is connected and is sufficient
(280–600 kPa (40 psi–87