6-61 
Failure 
Description 
Possible Cause  Recommended Action 
5.  Replace the internal AG 
module. 
6.  Replace the electronic 
vaporizer. 
00000040  The system exits the standby state.  During calibration, the system exits 
the standby state. 
00000080  The linearity does not meet the 
requirement. 
1.  Check whether the gas supply 
is sufficient. 
2.  Check whether the water trap 
and sampling line of the AG 
module are connected 
properly and whether leakage 
occurs. 
3.  Check whether leakage occurs 
in the internal pipeline of the 
machine so that the fresh gas 
cannot flow through the 
sampling point. 
4.  Replace the internal AG 
module. 
5.  Replace the electronic 
vaporizer. 
00000100  The system fails to write data into 
EEPROM. 
Replace the electronic vaporizer. 
00000200  The electronic vaporizer is 
automatically turned off due to an 
error. 
Perform diagnosis based on the 
fault code of the electronic 
vaporizer. 
00000400  The measured value of the AG module 
exceeds the upper limit during 
verification. 
1.  Calibrate the electronic 
vaporizer again. 
2.  Check whether the water trap 
and sampling line of the AG 
module are connected 
properly and whether leakage 
occurs. 
3.  Check whether leakage occurs 
in the internal pipeline of the 
machine so that the fresh gas 
cannot flow through the 
sampling point. 
4.  Replace the AG module. 
00000800  The measured value of the AG module 
exceeds the lower limit during 
verification. 
1.  Calibrate the electronic 
vaporizer again. 
2.  Check whether the water trap 
and sampling line of the AG