7-23 
VCM (refer to 8.6), and reconnect the 
cable between the backplane and display 
(refer to 8.1.7). 
4. If the problem persists, replace the 
monitoring and controlling assembly 
O2 Supply Failure  High 
The O2 supply pressure 
is low. 
1. Check whether gas is supplied 
normally. The gas supply pressure on the 
O2 pipeline should be within the range 
of 280- 600kPa (40-87PSI), and the 
cylinder pressure should be within the 
range of 6.9-20.0MPa (1000-2900PSI). 
Check the pressures displayed on the 
pressure gauge and status screen. 
2. If the pressure gauge and electronic 
pressure monitoring result indicate that 
the gas supply is sufficient, replace the 
supply pressure switch (refer to 11.2.1). 
Drive Gas Pressure 
Low 
High 
The drive gas pressure is 
low. 
1. Check whether gas is supplied 
normally. The gas supply pressure on the 
O2 pipeline should be within the range 
of 280- 600kPa (40-87PSI), and the 
cylinder pressure should be within the 
range of 6.9-20.0MPa (1000-2900PSI). 
Check the pressures displayed on the 
pressure gauge and status screen. 
2. If the gas supply connection is normal, 
replace the drive gas assembly (refer to 
ACGO 3-way 
Valve Failure 
Med 
status of the ACGO 
channel is inconsistent 
with the 3-way valve 
1. Replace the 3-way valve (refer to 
11.1.16). 
2. If the problem persists, replace the 
VCM (refer to 11.1.11). 
Auto Ventilation 
Disabled 
Low 
power-on self-test, and 
only manual ventilation 
Clear the power-on self-test alarm by 
referring to 7.3.1. 
Disabled - Leak 
Low 
The automatic circuit 
leak test fails. 
Handle the leak test failure by referring 
to 7.2.3. 
Aux Control 
Module Voltage 
Error 
Low 
3.3 V and 5 V supply 
voltages of the 
protective module are 
abnormal. 
1. Restart the anesthesia machine. 
2. If the power board has a voltage error 
alarm, clear this alarm (refer to 7.3.3). 
3. Select Setup > Service > Data 
Monitors > VPM and check whether the 
voltages in the A/D channel fall within 
the range specified in the A/D channel 
table. 
4. If the power supply voltage is out of 
the normal range, replace the monitoring 
and controlling assembly (refer to