7-30 
Cause  FRU  Solution 
00000002 
Boil-off space 
temperature sensor 
failure 
Electronic 
vaporizer 
1. Reinstall the electronic vaporizer, or 
restart the machine. 
2. If the problem persists, replace the 
00000004 
Heating Module 
Failure 
Electronic 
vaporizer 
1. Reinstall the electronic vaporizer, or 
restart the machine. 
2. If the problem persists, replace the 
00000008 
Drive gas pressure 
error in working 
state 
Electronic 
vaporizer base 
valve assembly 
Secondary 
pressure 
regulator 
Electronic 
vaporizer 
1. Replace the slot. If the alarm disappears, 
replace the vaporizer base valve assembly 
(refer to 
11.1.12). 
2. If the problem persists, insert another 
vaporizer. If the fault alarm disappears, 
replace the electronic vaporizer. 
3. If the problem persists, replace the 
pressure regulator of the electronic 
vaporizer (refer to 11.1.4). 
00000010 
Drive gas pressure 
error in standby state 
Electronic 
vaporizer base 
valve assembly 
Electronic 
vaporizer 
1. Replace the slot. If the alarm disappears, 
replace the vaporizer base valve assembly 
(refer to 
11.1.12). 
2. If the problem persists, insert another 
vaporizer. If the fault alarm disappears, 
replace the electronic vaporizer. 
00000020 
Failed to collect the 
drive gas pressure 
Replace the electronic vaporizer. 
00000040 
inconsistent with the 
Electronic 
vaporizer 
Replace the electronic vaporizer. 
00000080 
Valve supply voltage 
abnormal 
Replace the electronic vaporizer. 
00000100 
Injector supply 
voltage abnormal 
Replace the electronic vaporizer. 
00000200 
Analog 5V voltage 
abnormal 
Replace the electronic vaporizer. 
00000400 
Analog -5V voltage 
abnormal 
Replace the electronic vaporizer. 
00000800 
Replace the electronic vaporizer. 
00001000 
The injection times 
detected by the 
injection detection 
device are less than 
the actual injection 
Electronic 
vaporizer 
Replace the electronic vaporizer.