Loading...
11-16
No. Name P/N Test Item After
Replacement
Remarks
1
Airway pressure gauge
082-003308-00
5.2 System Check
/
2
Upper-left cover plate
FRU
115-066828-00
5.7.1 Vaporizer
Monitoring Function Test
The pressure
gauge is
removed from
the left cover
plate, and there
is no
auto/manual
switch.
3
Auto/manual switch
assembly
115-064076-00
5.2 System Check;
5.7.1 Vaporizer
Monitoring Function Test
/
4 I/O interface board 051-003189-00
Check whether the
network, USB, and DB9
ports are available.
/
5 O2 cell cable 009-009627-00
5.7.1 Vaporizer
Monitoring Function
Test
; Check whether the
O2 cell is properly
connected.
/
6
Work surface cover
plate FRU
115-066843-00
5.7.1 Vaporizer
Monitoring Function Test
Two waterproof
seals are
included.
7
Rubber plug for VE
door
049-001713-00 / /