11-24
No. Name P/N Test Item After
Replacement
Remarks
1
Pressure regulator
assembly (A9)
115-064092-00
5.7.1 Vaporizer
monitoring function test
This assembly is
applicable to A9.
It is used to
monitor the drive
gas pressure.
2
VCM FRU (with
ACGO/with auxiliary
pressure)for Canada
115-077992-00
6.3.1 Flow calibration
6.3.6 Pressure calibration
6.3.2 O2 sensor
calibration
5.2 System check
5.11 Active AGSS check
5.14
System ventilation
Performance
/
VCM FRU (with
ACGO/without
auxiliary pressure) for
US
115-077993-00
6.3.1 Flow calibration
6.3.6 Pressure calibration
6.3.2 O2 sensor
calibration
5.2 System check
5.11 Active AGSS check
5.14
System ventilation
Performance
/