Accutorr 7 Operator’s Manual 10 - 5
10.3 Disinfecting Temperature Probe
The recommended disinfectants include: ethanol 70%, isopropanol 70%,
glutaraldehyde-type 2% liquid disinfectants
To disinfect the temperature probe:
1. Disconnect the temperature probe from Temp connector.
2. Disinfect the probe with a soft cloth dampened with the recommended
disinfectant.
3. Wipe off all the remaining disinfectants from the probe with a soft cloth
dampened with water.
4. Dry the probe in a cool place.
WARNING
Properly dispose of the used soft cloth.