13 - 4 Accutorr 7 Operator’s Manual
13.5.4 Spot Check Trends Recording
The spot check trends strip includes patient information, recording time, measured
value and measurement complete time. When [Filter] is set to [Manually Saved] or [All],
the manual input data can be recorded.
To print the spot check trends,
1. Select the patient and filter.
2. Select [Print].
If several patients are selected, the recorder will print the data in sequence. Data from
different patients are separated by ”|”.
13.6 Removing a Paper Jam
If the recorder works incorrectly or produces unusual sound, check if there is a paper
jam. If a paper jam is detected, follow this procedure to remove it:
1. Open the recorder door.
2. Take out the paper and tear off the wrinkled or torn part.
3. Reload the paper and close the recorder door.
Refer to 13.2 Loading Paper for additional information.
13.7 Cleaning the Recorder Printhead
If the recorder has been used for a long time, deposits of paper debris may collect on the
printhead compromising the print quality and shortening the lifetime of the roller.
Follow this procedure to clean the printhead:
1. Take measures against the static electricity such as Disposable Wrist Strap for the
work.
2. Open the recorder door and take out the paper.
3. Gently wipe around the printhead using cotton swabs dampened with alcohol.
4. After the alcohol has completely been dried, reload the paper and close the
recorder door.
Refer to 13.2 Loading Paper for additional information.