6 - 8 Accutorr 7 Operator’s Manual
6.6 Pausing Alarms
You can temporarily disable alarm sound of all alarms by pressing and holding
hardkey on the panel for above 2 seconds.
When alarms are paused,
â– The alarms pause symbol and the remaining alarm pause time are displayed
in Alarm Information area.
â– For physiological alarms,
EtCO
2
0 to 32mmHg:
does not have
auto-
adjustable
alarm limit
0 to 32mmHg:
does not have
auto-
adjustable
alarm limit
0 to 32mmHg:
does not have
auto-
adjustable
alarm limit
0 to 32mmHg:
does not have
auto-
adjustable
alarm limit
Same as the
measurement
range
32 to 35mmHg:
29mmHg
32 to 35mmHg:
29mmHg
32 to 35mmHg:
41mmHg
32 to
35mmHg:
41mmHg
35 to 45mmHg:
(EtCO
2
-6)
mmHg
35 to 45mmHg:
(EtCO
2
-6)
mmHg
35 to 45mmHg:
(EtCO
2
+6)
mmHg
35 to
45mmHg:
(EtCO
2
+6)
mmHg
46 to
48mmHg:39
mmHg
46 to
48mmHg:39
mmHg
46 to
48mmHg:51
mmHg
46 to
48mmHg:51
mmHg
>48mmHg:
does not have
auto-
adjustable
alarm limit
>48mmHg:
does not have
auto-
adjustable
alarm limit
>48mmHg:
does not have
auto-
adjustable
alarm limit
>48mmHg:
does not have
auto-
adjustable
alarm limit
FiCO
2
N/A N/A Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement
range
awRR awRR × 0.5 or 6
rpm
(whichever is
greater)
(awRR – 10) or
30 rpm
(whichever is
greater)
awRR × 1.5 or
30 rpm
(whichever is
smaller)
(awRR+25) or
85 rpm
(whichever is
smaller)
Adult/pediatric:
6 to 55 rpm
Neonate: 10 to
90 rpm
Parameter
Low alarm Limit High alarm Limit
Auto-set alarm
limit Range
Adult/
Pediatric
Neonate
Adult/
Pediatric
Neonate