14 - 6 Accutorr 7 Operator’s Manual
14.2.3 EMR Communication Setup
1. Select [Main] → [Maintenance >>] → [User Settings >>] → Enter the required
password → Select [Ok] to access User Settings menu.
2. Select [Network>>] → [EMR Communication>>] to access [EMR
Communication Setup] menu.
3. Set the IP address/domain name and port.
4. Select [Ok] to exit the menu.
• The domain name supports only English characters, symbols and numbers.
14.2.4 NTP Server Setup
1. Select [Main] → [Maintenance >>] → [User Settings >>] → Enter the required
password → Select [Ok] to access User Settings menu.
2. Select [Network>>] → [NTP Setup>>].
3. Select [NTP Setup>>] to enter the NTP Setup menu. Set the following properties:
◆ Set IP address and port.
◆ Set the interval.
4. Select [Ok] to exit the menu.
To manually start one time synchronization with the NTP server, follow this procedure:
1. Select [Main] → [Maintenance >>] → [User Settings >>] → Enter the required
password → Select [Ok] to access User Settings menu.
2. Select [Network>>] → [NTP Setup>>] → [Manual].
14.2.5 Realtime Data Send Interval (for HL7 only)
The monitor will send out the monitored parameter data at the set intervals.
1. Select [Main] → [Maintenance >>] → [User Settings >>] → Enter the required
password → Select [Ok] to access User Settings Menu.
2. Select [Network>>] → [EMR Communication>>].
3. Set [Real Data Send] to an appropriate setting.
14.2.6 Alarm Server Settings (for HL7 only)
1. Select [Main] → [Maintenance >>] → [User Settings >>] → Enter the required
password → Select [Ok] to access User Settings menu.
2. Select [Network>>] → [Alarm Server Setup].
3. Set the IP address and port.