Accutorr 7 Operator’s Manual 16 - 5
16.4 Checking Monitor Information
Select [Main] → [Maintenance >>] → [Monitor Information>>], you can view
â–  system software version
â–  copyright information
â–  system configuration by selecting [Monitor Configuration>>], or
â–  status information, such as start time, self-test error, and so on by selecting
[Monitor Log>>]
You can print out the log information for the convenience of troubleshooting. Select
[Recorder] from the [Monitor Log] menu to do recording. The information will not be
saved after system shutdown.
16.5 Visual Inspection
Perform an overall inspection on the appearance of the equipment. The test is passed if
the equipment has no obvious signs of damage. Follow these guidelines when
inspecting the equipment:
1. Carefully inspect the case, display screen, buttons, and knob for obvious signs of
damage.
2. Inspect all external connections for loose connectors, bent pins or frayed cables.
3. Inspect all connectors on the equipment for loose connectors or bent pins.
4. Make sure that safety labels and data plates on the equipment are clearly legible.

Table of Contents

Other manuals for Mindray Accutorr 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr 7 and is the answer not in the manual?

Mindray Accutorr 7 Specifications

General IconGeneral
BrandMindray
ModelAccutorr 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals