Accutorr 7 Operator’s Manual A - 9
A.7.3 NIBP
Standards Meet standards of IEC80601-2-30, EN1060-1, EN1060-3, EN1060-4
and SP10
Technique Oscillometry
Mode of
operation
Manual, Auto, STAT and Program
Auto mode
repetition
intervals
1, 2, 2.5, 3, 5, 10, 15, 20, 30, 60, 90 or 120 min
STAT mode cycle
time
5 min
Max
measurement
time
Adult, pediatric: 180 s
Neonate: 90 s
Heart rate range 30 to 300 bpm
Measurement
ranges (mmHg)
Adult Pediatric Neonate
Systolic 25 to 290 25 to 240 25 to 140
Diastolic 10 to 250 10 to 200 10 to 115
Mean 15 to 260 15 to 215 15 to 125
Accuracy Max mean error: 5 mmHg
Max standard deviation: 8 mmHg
Static pressure
measurement
range
0 mmHg to 300 mmHg
Static pressure
measurement
accuracy
±3 mmHg
Resolution 1mmHg
Initial cuff
inflation
pressure range
(mmHg)
Adult:
Pediatric:
Neonate:
80 to 280
80 to 210
60 to 140