A - 10 Accutorr 7 Operator’s Manual
* Measurement accuracy verification: In adult and pediatric modes, the blood pressure
measurements measured with this device are in compliance with the American National
Standard for Electronic or Automated Sphymomanometers (ANSI/AAMI SP10) in terms
of mean error and standard deviation by comparing with intra-arterial or auscultatory
measurements (depending on the configuration) in a typical patient population. For
auscultatory reference, the 5th Korotkoff sound was used to determine the diastolic
pressure.
In neonatal mode, the blood pressure measurements measured with this device are in
compliance with the American National Standard for Electronic or Automated
Sphymomanometers (ANSI/AAMI SP10) in terms of mean error and standard deviation
by comparing with intra-arterial measurements (depending on the configuration) in a
typical patient population.
Default initial
cuff inflation
pressure
(mmHg)
Adult:
Pediatric:
Neonate:
160
140
90
Software
overpressure
protection
Adult:
Pediatric:
Neonate:
297±3 mmHg
297±3 mmHg
147±3 mmHg
Hardware
overpressure
protection
Adult:
Pediatric:
Neonate:
≤330 mmHg
≤330 mmHg
≤165 mmHg
Alarm limit Range (mmHg) Step (mmHg)
Sys High Adult: (low limit+5) to 290
Pediatric: (low limit+5) to 240
Neonate: (low limit+5) to 140
5
Sys Low 25 to (high limit-5)
Mean High Adult: (low limit+5) to 260
Pediatric: (low limit+5) to 215
Neonate: (low limit+5) to 125
Mean Low 15 to (high limit-5)
Dia High Adult: (low limit+5) to 250
Pediatric: (low limit+5) to 200
Neonate: (low limit+5) to 115
Dia Low 10 to (high limit-5)