Accutorr 7 Operator’s Manual B - 1
B EMC and Radio Regulatory Compliance
B.1 EMC
The device meets the requirements of IEC 60601-1-2. All the accessories listed in Chapter
also meet the requirements of IEC 60601-1-2 when in use with this device.
WARNING
• Using accessories, transducers and cables other than those specified may
result in increased electromagnetic emission or decreased electromagnetic
immunity of the patient monitoring equipment.
• The device or its components should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the device or its
components should be observed to verify normal operation in the
configuration in which it will be used.
• The device needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided
below.
• Other devices may interfere with this device even though they meet the
requirements of CISPR.
• When the inputted signal is below the minimum amplitude provided in
technical specifications, erroneous measurements could result.
• Portable and mobile communication equipment may affect the performance
of this monitor.
• Other devices that have RF transmitter or source may affect this device (e.g.
cell phones, PADs, PCs with wireless function).

Table of Contents

Other manuals for Mindray Accutorr 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr 7 and is the answer not in the manual?

Mindray Accutorr 7 Specifications

General IconGeneral
BrandMindray
ModelAccutorr 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals