D - 2 Accutorr 7 Operator’s Manual
D.2 Technical Alarm Messages
D.2.1 NIBP Alarm Messages
Alarm Message Clearable?
(Yes/No)
Alarm
Level
Cause Solution
NIBP-Sys Limit Err No Low The parameter
alarm limit is
inadvertently
changed.
Contact Mindray or
your service
personnel.
NIBP-Dia Limit Err No Low
NIBP-Mean Limit
Err
No Low
NIBP-Sys Over
Upper Limit
Yes Low The measured
pressure is greater
than the specified
NIBP measurement
upper limit.
Check the patient’s
condition and keep
the patient relaxed
and still. If the error
remains, contact
Mindray or your
service personnel.
NIBP-Dia Over
Upper Limit
Yes Low
NIBP-Mean Over
Upper Limit
Yes Low
NIBP-Sys Over
Lower Limit
Yes Low The measured
pressure is lower
than the specified
NIBP measurement
lower limit.
Restart the monitor
and retry. If the error
remains, contact
Mindray or your
service personnel.
NIBP-Dia Over
Lower Limit
Yes Low
NIBP-Mean Over
Lower Limit
Yes Low
NIBP SelfTest Err Yes High SelfTest Failed. The
cause may be the
transducer or A/D
sampling error.
Restart the monitor
and retry. If the error
remains, contact
Mindray or your
service personnel.
NIBP Init Err Yes Low An error occurred to
the NIBP module, or
there is a problem
with the
communications
between the
module and the
monitor.
Restart the monitor.
If the error remains,
contact Mindray or
your service
personnel.

Table of Contents

Other manuals for Mindray Accutorr 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr 7 and is the answer not in the manual?

Mindray Accutorr 7 Specifications

General IconGeneral
BrandMindray
ModelAccutorr 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals