Accutorr 7 Operator’s Manual D - 3
NIBP Comm Err Yes High An error occurred to
the NIBP module, or
there is a problem
with the
communications
between the
module and the
monitor.
Restart the monitor.
If the error remains,
contact Mindray or
your service
personnel.
NIBP Loose Cuff Yes Low The NIBP cuff is not
properly connected,
or there is a leak in
the airway.
Check the patient’s
condition and verify
patient type.
Replace with an
appropriate cuff and
connect it correctly.
If the problem still
exists, contact
Mindray or your
service personnel.
NIBP Air Pressure
Err
Yes Low An error occurred to
the air pressure.
Check the air
pressure. Restart the
monitor and retry. If
the error remains,
contact Mindray or
your service
personnel.
NIBP Weak Signal Yes Low The patient’s pulse
is weak or the cuff is
loose.
Check the patient’s
condition and
change the cuff
application site. If
the error persists,
replace the cuff.
NIBP Overrange Yes Low The measured NIBP
value is not within
the specified range.
Contact Mindray or
your service
personnel.
NIBP Excessive
Motion
Yes Low Patient’s arm moves
too much.
Check the patient’s
condition and
reduce the patient
motion.
NIBP Cuff
Overpress.
Yes Low The NIBP airway
may be occluded.
Check the airway
and measure again.
Alarm Message Clearable?
(Yes/No)
Alarm
Level
Cause Solution

Table of Contents

Other manuals for Mindray Accutorr 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr 7 and is the answer not in the manual?

Mindray Accutorr 7 Specifications

General IconGeneral
BrandMindray
ModelAccutorr 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals