D - 4 Accutorr 7 Operator’s Manual
NIBP Signal
Saturated
Yes Low The NIBP signal is
saturated.
Check the patient’s
condition and
reduce the patient
motion or other
sources.
NIBP Air Leak Yes Low The NIBP airway
may leak air.
The NIBP cuff is not
properly connected,
or there is a leak in
the airway.
NIBP Equip Err Yes High System error; or
pump, A/D
sampling or
pressure transducer
error; or pointer
error during
software running.
Check the patient’s
condition and NIBP
connections, or
replace the cuff.
NIBP Timeout Yes Low Time is out. In
Adult/Pediatric
mode, the
measurement time
is over 120 seconds;
in neonate mode,
the time is over 90
seconds.
NIBP Cuff Type
Wrong
Yes Low The cuff type
applied mismatches
the patient
category.
Verify the patient
category and
replace the cuff.
NIBP Illegally
Reset
Yes Low An illegal reset
occurred during
NIBP measurement.
Check if the airway
is occluded.
VeniPuncture
timeout
Yes Low System deflates the
cuff after a certain
time.
No operation is
required.
Alarm Message Clearable?
(Yes/No)
Alarm
Level
Cause Solution

Table of Contents

Other manuals for Mindray Accutorr 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr 7 and is the answer not in the manual?

Mindray Accutorr 7 Specifications

General IconGeneral
BrandMindray
ModelAccutorr 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals