Accutorr 7 Operator’s Manual C - 1
C Default Configurations
This chapter lists some of the most important factory default settings in configuration
management. You cannot change the factory default configuration itself. However, you
can make changes to the settings from the factory default configuration and then save
the changed configuration as a user configuration.
C.1 Parameter Configuration
C.1.1 NIBP
Name Default Config Saved at
Power Failure
Affected by
Defaults
Alarm On/Off On Yes Yes
Alarm Priority Med Yes Yes
Interval Manual Yes Yes
Clock On Yes Yes
NIBP End Tone 0 Yes Yes
Program Program 1 Yes Yes
Cuff Press.
(mmHg) in
VeniPuncture
Adult 80 Yes Yes
Pediatric 60
Neonate 40
Initial Pressure Adult 160 Yes No
Pediatric 140
Neonate 90
NIBP-Sys High
(mmHg)
Adult 160 Yes Yes
Pediatric 120
Neonate 90
NIBP-Sys Low
(mmHg)
Adult 90 Yes Yes
Pediatric 70
Neonate 40