Accutorr 7 Operator’s Manual 14 - 7
4. Select [Ok] to exit the menu.
14.2.7 Setting the Data Encryption Switch
The data encryption switch is off by default. To turn on the data encryption switch,
follow this procedure:
1. Select [Main] → [Maintenance >>] → [User Settings >>] → Enter the required
password → Select [Ok].
2. Select [Network>>].
3. Set [Data Encrypt Switch] to [On].
• Before switching on the data encryption switch, ensure that the eGateway
version you are using is version 6.4.0 or greater.Older version of the
eGateway may fail to handle the patient information.
14.3 Connecting the monitor to the CMS
To connect the monitor to the CMS, proceed as follows:
1. Select [Main] → [Maintenance >>] → [User Settings >>] → enter the required
password → Select [Ok] → [Network >>] → [Monitor Network Setup >>].
2. Set [Network Type] and [Address Type].
3. Input the monitor IP address, subnet mask and gateway address if the [Address
Type] is set to [Manual].
4. Connect the monitor to the CMS through either of the following methods:
◆ Admit the monitor on the CMS. Refer to the Hypervisor VI Operator’s Manual
(PN: 300B-20-47610) for details of admitting a monitor.
â—† Setting the CMS (refer to 14.3.1 Setting the CMS for details), and then
selecting a CMS (refer to 14.3.2 Selecting a CMS for details).
14.3.1 Setting the CMS
You can configure up to 30 central stations (CMS) for your monitor. To set the CMSs,
1. Select [Main] → [Maintenance>>] → [User Settings >>] → enter the required
password → Select [Ok] → [Network >>].
2. Set [Select CMS] to [On].
3. Select [Central Station Setup >>].
4. Set CMS names and corresponding IP addresses.