Accutorr 7 Operator’s Manual 6 - 9
alarm indicators are not shown. New physiological alarms will not be presented. If
the alarm condition still exists once the pause time expires, the alarms will be
presented.
For technical alarms, alarm sounds are paused, but alarm lamps and alarm
messages remain presented.
The default alarm pause time is 2 minutes. When the alarm pause time expires, or the
low battery alarm occurs, the alarm paused status is automatically deactivated. You can
also cancel the alarm paused status by pressing and holding the hardkey for more
than 2 seconds.
6.7 Switching Off Alarm Sound
When alarm volume is set to 0, the alarm sound is turned off. In the audio alarm off state,
The Audio Off symbol is displayed.
Audio indication of all alarms is suspended.
You can cancel the alarm sound off status by setting alarm volume to a value from 1 to
10.
WARNING
Pausing or switching off alarms may result in a hazard to the patient. Please
be very careful.
6.8 Switching Off [No CO2 Change Detected] Alarm
To switch off the [No CO2 Change Detected] alarm, follow this procedure:
1. Select [Main] [Maintenance >>] [User Settings >>] enter the required
password Select [Ok].
2. Select [Alarm Setup >>].
3. Set [No CO2 Change Detected Alarm Off] to [On].
4. Select [Main] [Alarm Setup >>].
5. Set [No CO2 Change Detected] to [Off].
.
WARNING
Switching off [No CO2 Change Detected] alarm may result in a hazard to the
patient. Always keep the patient under close surveillance.
NOTE