Accutorr 7 Operator’s Manual 9 - 9
9.9.3 Disinfection
Disinfect the cuff with a cloth dampened in 70% isopropanol or with ultraviolet light.
Also disinfect the bladder with ultraviolet light.
NOTE
• Prolonged use of disinfectant may cause discoloration of the cuff.
Replace the bladder after cleaning and disinfecting the cuff:
1. Place the bladder on the top of the cuff (as shown below).
2. Roll the bladder lengthwise and insert it into the large opening.
3. Hold the hose and the cuff and shake the complete cuff until the bladder is in
position.
4. Thread the hose from inside the cuff, and out through the small hole under the
internal flap.
9.9.4 Reusable Bladderless Cuffs
Clean cuffs with warm water and a mild detergent. Do not use a detergent containing
hand conditioners, softeners, or fragrances.
Antimicrobial Definition
Bladderless cuffs are treated with an antimicrobial coating. Antimicrobial technology
effectively controls a broad spectrum of bacteria, fungi, algae and yeasts on a wide
variety of treated substrates.

Table of Contents

Other manuals for Mindray Accutorr 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr 7 and is the answer not in the manual?

Mindray Accutorr 7 Specifications

General IconGeneral
BrandMindray
ModelAccutorr 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals