Accutorr 7 Operator’s Manual A - 1
A Product Specifications
A.1 Classifications
The equipment is classified, according to IEC60601-1:
A.2 Environmental Specifications
Main Unit
Type of protection against
electrical shock
CLASS I EQUIPMENT, equipment energized from an
external and internal electrical power source.
Degree of protection against
electrical shock
DEFIBRILLATION-PROOF TYPE CF AAPPLIED PART for
SpO
2
, NIBP and TEMP
DEFIBRILLATION-PROOF TYPE BF AAPPLIED PART for
CO
2
Mode of operation Continuous
Degree of protection against
harmful ingress of water
IPX1 (protection against vertically falling water drops)
degree of safety of application
in the presence of a
FLAMMABLE
ANAESTHETIC MIXTURE WITH
AIR or WITH OXYGEN OR
NITROUS OXIDE
EQUIPMENT not suitable for use in the presence of a
FLAMMABLE ANAESTHETIC MIXTURE WITH AIR or WITH
OXYGEN OR NITROUS OXIDE
Degree of mobility Portable
Item Operating conditions Storage conditions
Temperature (ºC) 0 to 40
5 to 40 (with Temp module)
10 to 40 (with CO
2
module)
-20 to 60
Relative humidity
(noncondensing)
15% to 95% 10% to 95%
Barometric (kPa) 57.0 to 107.4 16.0 to 107.4