Accutorr 7 Operator’s Manual 14 - 5
The recommended distance between the patient monitor and other non-Wi-
Fi wireless devices, including wireless devices at the frequency of 2.4GHz (e.g.
cellular mobile communication networks, microwave ovens, interphones,
cordless phones and electro-surgical units) is no less than 20 cm.
14.2 Setting Data Send Method
You can choose to send the monitor data to hospital system through eGateway or HL7
protocol.
1. Select [Main] [Maintenance >>] [User Settings >>] Enter the required
password Select [Ok] [Network>>].
2. Set [Data Send Method] to [eGateway] or [HL7].
14.2.1 DNS Setup
1. Select [Main] [Maintenance >>] [User Settings >>] Enter the required
password Select [Ok] to access [User Settings] menu.
2. Select [Network>>] [Monitor Network Setup>>] [DNS Setup].
3. Set [Address Type], [Preferred DNS Server], and [Alternate DNS Server].
14.2.2 ADT Communication Setup
1. Select [Main] [Maintenance >>] [User Settings >>] Enter the required
password Select [Ok] to access [User Settings] menu.
2. Select [Network>>] [ADT Communication>>] to access [ADT
Communication Setup] menu.
3. Make the following settings:
Input an IP address or a domain name as the server address.
Input the port.
Switch on/off ADT query.
Set [Query from Local Facility]. If you set [Query from Local Facility] to
[Ye s ], you can query patient information only from local facility; If you set
[Query from Local Facility] to [No], you can query patient information from
all networked facilities.
4. Select [Ok] to exit the menu.
NOTE
The domain name supports only English characters, symbols and numbers.