EasyManuals Logo

Mindray Accutorr 7 Operator's Manual

Mindray Accutorr 7
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #161 background imageLoading...
Page #161 background image
Accutorr 7 Operator’s Manual A - 11
A.7.4 Temp
Statistical Results of Clinical Investigation Data
A.8 CO
2
Standards Meets standards of ASTM E1112, ASTM E1104, ISO 80601-2-
56
Technique Thermal resistance (use thermistor to measure
temperature)
Measurement range Monitor mode:25 to 44°C (77 to 111.2 °F)
Predictive mode: 35 to 43°C (95 to 109.4 °F)
Accuracy (Monitor mode) 25°C to 32°C (not include 32°C): ± 0.2°C
32°C to 44 °C (include 32°C): ±0.1°C
or
77°F to 89.6°F (not include 89.6°F): ± 0.4°F
89.6°F to 111.2°F (include 89.6°F): ± 0.2°F
Resolution ±0.1°C
Minimum measurement
time for accurate readings
Monitor mode: <60 s
Predictive mode: <20 s (typical test: < 12s )
Alarm limit Range Step
Temp High (low limit +1) °C to 44°C
(low limit +1.8) °C to 111.2°F
0.1°C
0.2°F
Temp Low 25°C to (high limit -1) °C
77°F to (high limit -1.8) °F
Clinical BIAS (Δ cb) Limits of Agreement
(LA)
Clinical
Repeatability (σr)
Oral 0.02 0.33 0.1
Axilla 0.06 0.38 0.13
Rectum -0.05 0.48 0.14
Standard Meets standard of ISO 80601-2-55
Technique Infrared absorption
[No CO2 Changes
Detected] delay time
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Measurement range 0 to 20% (0 to 150 mmHg)

Table of Contents

Other manuals for Mindray Accutorr 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr 7 and is the answer not in the manual?

Mindray Accutorr 7 Specifications

General IconGeneral
BrandMindray
ModelAccutorr 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals