6 - 10 Accutorr 7 Operator’s Manual
• If the alarm condition for [NO CO2 Change Detected] is triggered, the alarm
message will indicated "Check Breathing. No CO2 Changes Detected".
6.9 Resetting Alarms
By pressing the hardkey, you can reset the alarm system to acknowledging the on-
going alarms and enable the alarm system to respond to a subsequent alarm condition.
For physiological alarms, except the NIBP-related alarms, when the alarm system is reset:
â–  The alarm sound is silenced.
■ A √ appears before the alarm message, indicating that the alarm is
acknowledged.
â–  The icon appears in the alarm symbol area.
â–  The parameter numeric and alarm limits still flash.
The indication of alarm lamp for the physiological alarm depends on the alarm light
setting.
â–  When [Alarm Light on Alarm Reset] is set to [On], the alarm lamp remains
flashing.
â–  When [Alarm Light on Alarm Reset] is set to [Off], the alarm lamp stops flashing.
To set [Alarm Light on Alarm Reset]:
1. Select [Main] → [Maintenance >>] → [User Settings >>] → enter the required
password → Select [Ok] to access [User Settings] menu.
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Select [Alarm Light on Alarm Reset], and toggle between [On] and [Off].
The default setting for [Alarm Light on Alarm Reset] is [On].
Technical alarms give different alarm indicators when the alarm system is reset:
■ For some technical alarms, including the NIBP-related alarms, a √ appears before
the alarm message and appears in the alarm symbol area, indicating that the
alarm is acknowledged. The indication of the alarm lamp depends on the alarm
light setting.
â–  Some technical alarms are changed to the prompt messages.
â–  Some technical alarms are cleared. The monitor gives no alarm indications.
For details about the indications of technical alarms when the alarm system is reset,
refer to
D.2 Technical Alarm Messages.

Table of Contents

Other manuals for Mindray Accutorr 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr 7 and is the answer not in the manual?

Mindray Accutorr 7 Specifications

General IconGeneral
BrandMindray
ModelAccutorr 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals