11 - 2 Accutorr 7 Operator’s Manual
11.3 Measurement Limitations
The following factors may influence the measurement accuracy:
â–  Leaks or internal venting of sampled gas
â–  Mechanical shock
â–  Cyclic pressure up to 10 kPa (100 cmH
2
O)
â–  Other sources of interference, if any
Measurement accuracy may be affected by the breath rate and inspiration/expiration
ratio (I: E ratio) as follows:
â–  EtCO
2
value is within specification for breath rate ≤ 60 bpm and I/E ratio ≤ 1:1;
â–  EtCO
2
value is within specification for breath rate ≤ 30 bpm and I/E ratio ≤ 2:1.
Measurement accuracy is unspecified for breath rate larger than 60 bpm.
11.4 CO
2
Display
The CO
2
parameter and waveform areas provide FiCO
2
measurement, EtCO
2
measurement, awRR measurement, and a CO
2
waveform.
1. End tidal CO
2
(EtCO
2
)
2. Fraction of inspired CO
2
(FiCO
2
)
3. Airway respiration rate (RR)
4. CO
2
waveform
5. Scale
11.5 Automatic CO
2
Module Zeroing
The CO
2
module performs zero calibration automatically when needed.
NOTE
• The CO
2
module temporally stops measuring during zeroing.
1 2 3
4
5

Table of Contents

Other manuals for Mindray Accutorr 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr 7 and is the answer not in the manual?

Mindray Accutorr 7 Specifications

General IconGeneral
BrandMindray
ModelAccutorr 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals