Accutorr 7 Operator’s Manual D - 7
D.2.3 Temperature Alarm Messages
Alarm Message Clearable?
(Yes/No)
Alarm
Level
Cause Solution
Temp Init Err Yes High An error occurred to
the Temp module,
or there is a
problem with the
communications
between the
module and the
monitor, or Temp
calibration error.
Restart the monitor.
If the error remains,
contact Mindray or
your service
personnel.
Temp Comm Err No High An error occurred to
the Temp module,
or there is a
problem with the
communications
between the
module and the
monitor.
Restart the monitor.
If the error remains,
contact Mindray or
service personnel.
Temp Alarm Limit
Err
No High The alarm limit of
Temp is changed
inadvertently.
Contact Mindray or
your service
personnel.
Warmup Timed
Out
Yes Med The initial probe
temperature in
measurement is too
high.
Cool the probe and
retry.
Warming Resistor
Err
No Med The thermal resistor
on the temperature
probe has an error
(Can not work
properly).
Replace
temperature probe.
Ambient Temp
Overrange
Yes Med The environmental
temperature is out
the range of the
monitor’s
measurement.
Change an
environment and
retry.
Temp Voltage Err Yes Med The voltage is too
high or too low.
Check the power
supply.
Temp Prediction
Err
Yes Low The measuring
operation is
improper.
Retry the
measurement.

Table of Contents

Other manuals for Mindray Accutorr 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr 7 and is the answer not in the manual?

Mindray Accutorr 7 Specifications

General IconGeneral
BrandMindray
ModelAccutorr 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals