Accutorr 7 Operator’s Manual 6 - 7
NIBP-S (SYS×0.68)
10 mmHg
(SYS-15) or 45
mmHg,
whichever is
greater
(SYS×0.86)
38 mmHg
(SYS+15) or
105 mmHg,
whichever is
smaller
Adult:
45 to 270
mmHg
Pediatric:
45 to 185
mmHg
Neonate:
35 to 115
mmHg
NIBP-D (Dia×0.68)6
mmHg
(Dia-15) or 20
mmHg,
whichever is
greater
(Dia×0.86)32
mmHg
(Dia15) or
80 mmHg,
whichever is
smaller
Adult:
25 to 225
mmHg
Pediatric:
25 to 150
mmHg
Neonate:
20 to 90 mmHg
NIBP-M (Mean×0.68)
8 mmHg
(Mean-15) or
35 mmHg,
whichever is
greater
(Mean×0.86)
35 mmHg
(Mean15)
or 95 mmHg,
whichever is
smaller
Adult:
30 to 245
mmHg
Pediatric:
30 to 180
mmHg
Neonate:
25 to 105
mmHg
Temp ( T-0.5) °C (T-0.5) °C (T+0.5) °C (T+0.5) °C Same as the
measurement
range
Parameter
Low alarm Limit High alarm Limit
Auto-set alarm
limit Range
Adult/
Pediatric
Neonate
Adult/
Pediatric
Neonate