Accutorr 7 Operator’s Manual B - 3
Guidance and Declaration — Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.
Immunity
Test
IEC 60601 Test
Level
Compliance Level Electromagnetic
Environment - Guidance
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
Transient/
burst IEC
61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge IEC
61000-4-5
± 1 kV lines(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV lines(s) to
line(s)
± 2 kV line(s) to
earth
Voltage dips,
Short
interruptions
and voltage
variation on
power supply
input lines IEC
61000-4-11
<5% U
T
(>95% dip
in U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip
in U
T
) for 5 seconds
<5% U
T
(>95% dip
in U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip
in U
T
) for 5 seconds
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
our product requires
continued operation during
power mains interruptions,
it is recommended that our
product be powered from
an uninterruptible power
supply or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
our product requires
continued operation during
power mains interruptions,
it is recommended that our
product be powered from
an uninterruptible power
supply or a battery.
Note: U
T
is the A.C. mains voltage prior to application of the test level.

Table of Contents

Other manuals for Mindray Accutorr 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr 7 and is the answer not in the manual?

Mindray Accutorr 7 Specifications

General IconGeneral
BrandMindray
ModelAccutorr 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals