Accutorr 7 Operator’s Manual D - 9
XX represents a measurement or parameter label.
CO2 SelfTest
Err
No High An error occured
during CO2
module
initialization.
Replace the
module.
CO2 Sensor
High Temp
No Low Ambient
temperature is
too high or there
is a module
failure.
Lower the
operating
temperature.
If the alarm
persists, the CO2
module may
have a failure.
Contact Mindray
or your service
personnel.
CO2 Zero
Failed
No Low The zeroing
failed.
Check the CO2
connections. If
the alarm
persists, contact
Mindray or your
service
personnel.
CO2 No
Filterline
Yes Low The sample line is
off or has been
disconnected.
Make sure that
the sample line is
connected.
CO2 FilterLine
Occluded
No Low The airway or
water trap is
occluded.
Check the airway
and remove the
occlusion.
XX Overrange No Low The measured
parameter or
measurement XX
value is not
within the
specified range.
Contact Mindray
or your service
personnel.
Alarm Message Clearable?
(Yes/No)
Alarm
Level
Cause Solution

Table of Contents

Other manuals for Mindray Accutorr 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Accutorr 7 and is the answer not in the manual?

Mindray Accutorr 7 Specifications

General IconGeneral
BrandMindray
ModelAccutorr 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals