A-7 
0% to 69%: Not specified. 
Refreshing rate  1 s 
Response time 
≤ 30 s (PI > 0.3, no disturbance, SpO2 value sudden change 
within 70% - 100%) 
* When the SpO
2 
sensor is applied for neonatal patients as indicated, the specified accuracy 
range is increased by ±1%, to compensate for the theoretical effect on oximeter 
measurements of fetal hemoglobin in neonatal blood. 
 
Information of the Test Subjects of the Clinical Study Report 
 
Skin color  Gender  Number  Age (years)  Health 
Black Male  1  28.2±9.19  Healthy 
Female 1 
Yellow Male  3 
Female 9 
 
 
A.6.2 PR 
Module  PR High Limit  PR Low Limit  Step (bpm) 
Mindray SpO
2
 Module  (low limit +2) to 254  20 to (high limit-2) 
1 
Masimo SpO
2
 Module  (low limit +2) to 240  25 to (high limit-2) 
Nellcor SpO
2
 Module  (low limit +2) to 300  20 to (high limit-2) 
NIBP Module  (low limit +2) to 240  40 to (high limit-2) 
 
 
PR from Mindray SpO
2
 Module 
Measurement range  20 to 254 bpm 
Resolution 1 bpm 
Response time 
≤ 30 s (PI > 0.3, no disturbance, PR value sudden change within 
25 --- 250 bpm) 
Accuracy  ±3 bpm (without motion) 
Refreshing rate  1 s 
 
 
PR from Masimo SpO
2
 Module 
Measurement range  25 to 240 bpm 
Resolution 1 bpm