A-8
Response time ≤ 20 s (PR value sudden change within 25 --- 240 bpm)
Accuracy
±3 bpm (without motion)
±5 bpm (with motion)
Refreshing rate 1 s
PR from Nellcor SpO
2
Module
Measurement range 20 to 300 bpm
Resolution 1 bpm
Response time
≤ 30 s (PI > 0.3, no disturbance, PR value sudden change within
25 --- 250 bpm)
Accuracy
20 to 250 bpm: ±3 bpm
251 to 300 bpm, not specified
Refreshing rate 1 s
PR from NIBP Module
Measurement range 40 to 240 bpm
Resolution 1 bpm
Accuracy ±3bpm or ±3%, whichever is greater
A.6.3 NIBP
Standards Meet standards of IEC 80601-2-30, ISO 81060-2
Technique Oscillometry
Mode of operation Manual, Auto, STAT and Program
Auto mode repetition
intervals
1, 2, 2.5, 3, 5, 10, 15, 20, 30, 60, 90 or 120 min
STAT mode cycle time 5 min
Max measurement time
Adult, pediatric: 180 s
Neonate: 90 s
Heart rate range 40 to 240 bpm
Measurement ranges
(mmHg)
Adult Pediatric Neonate
Systolic: 40 to 270 40 to 200 40 to 135
Diastolic: 10 to 210 10 to 150 10 to 100
Mean: 20 to 230 20 to 165 20 to 110