Error Code
Subsystem
Message for Error log
User Error Message
Action on
Action
0xa6030401
CBKEY30A
SMBUS Error
"IOE30A SMBUS
ERROR"
Continue
0xa6040401
CBKEY30A
Init Error
"IOE30A INIT error"
Continue
0xa6040402
CBKEY30A
Batt Status Retrieval Error
"IOE30A BATT STATUS
RETRIEVAL ERROR"
Continue
0xaf034005
HSSL_LINK
ostr.str
"Cart Device Initialization
Error. If error persists call
service."
Continue
0xaf050401
THERMALSER
VICES
Fan is stuck
"System fan Error:
Please shutdown
system."
Stop
0xaf050403
THERMALSER
VICES
System Overheat Warning
"System Overheat
Warning: Finish exam and
shutdown."
Continue
0xaf050405
THERMALSER
VICES
Battery Overheat Warning
"Battery Overheat
Warning: finish exam -
replace battery."
Continue
0xaf050430
THERMALSER
VICES
PWR_HV over warning
Temperature Limit
"System Overheat
Warning: Finish exam and
shutdown."
Continue
0xaf050431
THERMALSER
VICES
PWR_LV over warning
Temperature Limit
"System Overheat
Warning: Finish exam and
shutdown."
Continue
0xaf050432
THERMALSER
VICES
PWR_TSUPR over warning
Temperature Limit
"System Overheat
Warning: Finish exam and
shutdown."
Continue
0xaf050433
THERMALSER
VICES
DIG_HYDRA over warning
Temperature Limit
"System Overheat
Warning: Finish exam and
shutdown."
Continue
0xaf050434
THERMALSER
VICES
DIG_VSUPR over warning
Temperature Limit
"System Overheat
Warning: Finish exam and
shutdown."
Continue
0xaf050435
THERMALSER
VICES
XMT_V5NA over warning
Temperature Limit
"System Overheat
Warning: Finish exam and
shutdown."
Continue
0xaf050436
THERMALSER
VICES
XMT_VSUPR over warning
Temperature Limit
"System Overheat
Warning: Finish exam and
shutdown."
Continue
0xaf050438
THERMALSER
VICES
RCV_AFE over warning
Temperature Limit
"System Overheat
Warning: Finish exam and
shutdown."
Continue
0xaf050439
THERMALSER
VICES
RCV_CW over warning
Temperature Limit
"System Overheat
Warning: Finish exam and
shutdown."
Continue
0xaf05043a
THERMALSER
VICES
RCV_LV over warning
Temperature Limit
"System Overheat
Warning: Finish exam and
shutdown."
Continue

Table of Contents

Other manuals for Mindray Zonare ZS3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Zonare ZS3 and is the answer not in the manual?

Mindray Zonare ZS3 Specifications

General IconGeneral
BrandMindray
ModelZonare ZS3
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals