ZS3 Service Manual
Page 96 of 295
Name
NOTE
Cancel
Click to cancel the parameter setting.
Update
Select an item in the service list, change the
parameters in the above area, and click [Update] to
update the item in the service list.
Service
list
Delete
Click to delete the selected service in the service list
Default
Set the Worklist server as the default.
Select an item in the service list, click [Default] and
you can see Y in the Default column.
Verify
Click to verify if the two DICOM application entities are
normally connected.
Exit
Click to exit the screen.
NOTE:
Set the MPPS service as the default when using the MPPS.
STORAGE COMMITMENT SETTING
1. Select [Setup][DICOM][DICOM Service Setting][Storage] to enter the storage
page.
2. Select the device name, and type the information about the AE title, the port.
3. Click [Add]. The entry will be added in the service list.