4.4.7.2 Connecting a load cell with a 4-wire cable
The following links between the terminal contacts are provided:
from SIGN. SENSE S+ after LC SUPPLY V+
from SIGN. SENSE S- after LC SUPPLY V-
Terminal Port/color code Description
SIGN. M+ + Meas./gn + Measuring voltage (load cell output)
SIGN. M- - Meas./gy - Measuring voltage (load cell output)
SENSE S+ + sense + Sense voltage
SENSE S- - Sense - Sense voltage
LC SUPPLY V+ + Supply/rd + Supply voltage
LC SUPPLY V- - Supply/bu - Supply voltage
LC SUPPLY
Grounding Screen (ground)
4 Device installation Transmitter Series PR 5220
EN-39 Minebea Intec