Dansensor® Lippke® 5000 EN User Guide 125
P/N 380419-C
05/2022
Basic specifications
Measuring range 1 - 5000 mbar
Displayed resolution 0 - 1000 mbar: 0,1 mbar
1000 - 5000 mbar: 1 mbar
Accuracy 1 - 350 mbar: ± 0.5 mbar or ±0.5% of reading
350 - 5000 mbar: ± 5 mbar or ±0.5% of reading
Repeatability 1 - 350 mbar:
Regulation at a fixed pressure: ± 0.5 mbar
Overshoot regulation to a fixed pressure:
≤ 1.5 mbar + 0.25% of set-point
350 - 5000 mbar:
Regulation at a fixed pressure: ± 3 mbar
Overshoot regulation to a fixed pressure:
≤ 5 mbar + 0.2% of set-point
Pressure units mbar, mmHG or psi
Test time 1 - 500 sec.
Calibration interval 12 months
Data Collection
Number of test 300
definitions
Data collection Test data from more than 1.000.000 tests
capacity
Collection type Stand alone units: Exported to CSV file
Using PC Software: Local or external SQL database
Conformity
CE mark
China RoHS II
Other standards
ASTM F-1140, F-2054, F-2095, F-2096
CFR 21 part 11 (with PC SW or use label printer, stand-alone without printer not approved).
ISO 11607