Dansensor® Lippke® 5000 EN User Guide 79
P/N 380419-C
05/2022
Error messages
NOTE! The error message list below is not complete but describes the possible
errors that the daily user has an opportunity to respond to and correct.
Other error messages may occur, but these will most often require a greater
technical insight into the device or that you contact your assigned service provider
(see
"Service" on page 108).
NOTE! If problems or errors occur, that are not described in this section or if one or
more errors persist, you should contact your assigned service provider (see
"Service" on page 108).
No. Message Problem Solution
3004 Name exists: XX This name is already used. Use another name.
3005 ID already exist: XX This ID is already used. Use another ID.
3009 Binary file version not
supported
The binary file is not
compatible with the
device.
Check that it was the
correct file. Files from
other Dansensor
equipment can not be
imported.
3010 Printing failed The printer was not
connected or not turned
on.
Non-compatible printer is
used.
Connect and turn on the
printer.
Change to recommended
printer.
3011 Max. pressure < prefill
pressure
Incorrect parameter
settings when creating a
Burst Test Definition.
Set appropriate values.
3012 Burst minimum limit is
lower than prefill pressure
Incorrect parameter
settings when creating a
Burst Test Definition.
Set appropriate values.
3013 Max. burst pressure is
lower than max. burst limit
Incorrect parameter
settings when creating a
Burst Test Definition.
Set appropriate values.
3014 Burst maximum limit is
lower than burst
minimum limit
Incorrect parameter
settings when creating a
Burst Test Definition.
Set appropriate values.
3015 Pressure drop limit is
greater or equal with test
pressure
Incorrect parameter
settings when creating a
Leak Test Definition.
Set appropriate values.
3016 Lower time limit is greater
or equal with upper time
limit
Incorrect parameter
settings when creating a
Creep2Fail Test Definition
Set appropriate values.