Dansensor® Lippke® 5000 EN User Guide 131
P/N 380419-C
05/2022
Operating/storing temperatures
The device is designed to be operated and stored in the temperatures ranges as specified in
"Basic specifications" on page 125.
The device can be damaged if stored or operated outside of these temperature ranges.
Avoid exposing the device to dramatic changes in temperature or humidity.
NOTE! If the device has been stored in a cold location and then moved to a warmer
location then at least one hour of acclimatization is required before switching on
the unit.
If operated in a cold and high humidity environment it is advisable to let the device
heat up for 30 minutes after is has been switched on.
NOTE! The device must be placed in a temperature controlled environment in order
to perform within specifications. Temperature and pressure in a (closed) test
volume are proportional i.e. changes in ambient temperature will have an impact
on the pressure during a leak test.