Dansensor® Lippke® 5000 EN User Guide 53
P/N 380419-C
05/2022
3. Setting Up
Compressed air connections
Compressed air supply requirements
The compressed air supply must be dry, clean and free from oil - see specifications in
"Compressed air supply" on page 124.
We recommend to use our optional pressure regulator/filter unit which ensures that the air
quality fulfills the required specifications - see
"Pressure Regulator with filters" on page 30 for
details.
NOTE! The above regulator/filter unit does not ensure that the air is dry and
therefore it is recommended to install a type of air-drying unit as well.
NOTE! If the air supply pressure is below or above the specified input pressure, the
system may not function as intended.
CAUTION! If the specified maximum pressure is exceeded, there is a danger of
bursting of tubing and other components.

Table of Contents