EasyManua.ls Logo

Mocon AMETEK Dansensor Lippke 5000 - Pressure Regulator with Filters; Package Pressure Release Valve

Mocon AMETEK Dansensor Lippke 5000
136 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30 User Guide EN Dansensor® Lippke® 5000
P/N 380419-C
05/2022
Pressure Regulator with filters
The optional pressure regulator/filter unit ensures that the supplied compressed air quality is
compliant with DIN ISO 8573-1:2010 [4:4:3].
Package pressure release valve
When testing packages with very large volumes, it can be very time consuming to ventilate
the package through the device after the test. In these cases, we recommend using the
pressure release valve, which must be fitted in the “FEED” line between the package and the
appliance.
The valve also makes it possible to ventilate packages that do not burst in connection with a
“Burst”- or “Creep” test.
The valve is fitted with a noise filter.
NOTE! When using the pressure release valve, the “Ventilation” parameter must be
set to “Off” or “Manual” in the appropriate test definition.

Table of Contents

Related product manuals